A high oxygen reading on a natural gas stream usually means air entered the process, air entered the sample path, liquid water temporarily offset the reading, or the analyzer range and calibration no longer match the process.
Work the process first, then the sample path, then the analyzer.
Air can enter at tank hatches, VRU suction, compressor rod packing or shaft seals when suction is low, wellheads near atmospheric pressure, or gathering fittings. Recent maintenance that opened piping to air can leave a high reading until the system is purged. Purge or blanketing flow that is too low on tanks or flare headers can also admit air.
EPA Natural Gas STAR guidance on installing vapor recovery units on storage tanks describes a control pilot that shuts the compressor down and lets vapor flow back into the tank, so a vacuum does not form when the oil level drops. That guidance does not measure oxygen in the vapor line.
Leaks on the sample line can pull air toward the Sensor Module. Liquid water in the sample path offsets the reading until the path is purged. OxyScout™ is specified for non-condensing service. Air left in tubing after calibration or service can also read high until displaced.
Confirm the installed model and range: Ultra-Trace (OxyScout-UT, 0–50 ppm), Trace (OxyScout-T, 0–1,000 ppm), or Percent (OxyScout-P). Confirm zero and span practice. Two-point calibration is recommended. A one-point calibration is acceptable. Air doesn't damage or poison the OxyScout optical sensor. Once process gas is flowing again, the reading typically recovers within minutes. Electrochemical trace O₂ cells can be used up or permanently damaged by extended air exposure.
The arithmetic below is hypothetical. NOAA lists the oxygen content of dry air, excluding water vapor, as 20.946 percent. A hypothetical mix of oxygen-free gas with 100 ppm by volume of that dry air adds 20.946 ppm of oxygen. The figure is not a field measurement and not an OxyScout specification.
| Hypothetical mix | Oxygen added |
|---|---|
| 100 ppm by volume of dry air in oxygen-free gas | 20.946 ppm |
100 ppm × 0.20946 = 20.946 ppm.
For range selection against a tariff limit, see Pipeline Oxygen Limits From 10 ppm to 0.2%. For a quote, open the quote page.
A high oxygen reading on a natural gas stream usually means air entered the process, air entered the sample path, liquid water temporarily offset the reading, or the analyzer range and calibration no longer match the process. Work the process first, then the sample path, then the analyzer.
Liquid water in the sample path offsets the reading until the path is purged. OxyScout is specified for non-condensing service.
Ultra-Trace (OxyScout-UT, 0–50 ppm). Trace detection is 50 ppm, so a 10 ppm tariff example requires Ultra-Trace.
Air doesn't damage or poison the OxyScout optical sensor. Once process gas is flowing again, the reading typically recovers within minutes. Electrochemical trace O₂ cells can be used up or permanently damaged by extended air exposure.