Calibration keeps a continuous oxygen analyzer aligned with known gases. For OxyScout™, a two-point calibration with zero gas and span gas is recommended. A one-point calibration is acceptable.
The International Vocabulary of Metrology defines calibration as an operation that, under specified conditions, establishes a relation between quantity values from measurement standards and the corresponding indications of the instrument. The same entry says calibration should not be confused with adjustment of a measuring system.
For OxyScout, a two-point calibration with zero gas and span gas is recommended. A one-point calibration is acceptable. The steps below are the OxyScout zero and span procedure from the user manual.
OxyScout uses a two-point calibration: a zero on oxygen-free gas, then a span on a gas with a known upper point. ScoutView PC software, included with every analyzer, walks through guided zero and span. You can also trigger the same calibrations from an RTU, PLC, or laptop Modbus master. Either method gives the same result. The gas-handling steps below apply to both.
A one-point calibration, often a zero calibration with pure nitrogen, is also acceptable. This is sometimes the path chosen for custody transfer or other applications where any oxygen reading above 0 ppm is unacceptable. This often saves users time and money, since a span gas is not needed.
Calibrate at initial startup, after every Sensor Module replacement, and periodically after that. A typical interval for most installations is 1–2 months. Drift between calibrations also depends on the sample rate and on process quality, as described later on this page.
Zero gas is ultra-high-purity (UHP) nitrogen, 99.999%, certified to less than 1 ppm O₂. Use a stainless-steel-diaphragm regulator. An elastomer-diaphragm regulator will contaminate the gas with atmospheric oxygen and make a true zero impossible.
Span gas should be a certified mixture near your operating range or critical point. For a 10 ppm application, that is typically 10 ppm O₂ in nitrogen. Treat span gas with the same care as zero gas: a stainless-diaphragm regulator, bled thoroughly, stainless tubing, and a leak check. If a span gas reads far from its certificate value, suspect the gas handling or the cylinder before suspecting the analyzer. An air calibration is often more accurate than a mishandled low-level span gas.
A newly installed regulator is full of air on both sides of its diaphragm. Simply flowing gas through it can take days to clear. Bleed the regulator with the steps below before trusting any gas that passes through it.
Always calibrate with the analyzer venting as it will run in service: the same vent path and the same constant flow setting. Calibrating at one vent condition and operating at another builds a permanent offset into every reading.
Vent to atmosphere is recommended. A return to a process point is acceptable if that point stays at a steady pressure lower than the regulated supply, so gas keeps flowing and the calibration holds.
Do this for the zero-gas regulator and, later, for the span-gas regulator.
Drift between calibrations depends on the sample rate set and on process quality. Typically, a 1–2 month interval results in low single-digit ppm of oxygen drift across OxyScout models. That statement describes drift between calibrations. Drift is not the same as detection limit or accuracy.
Ultra-Trace OxyScout-UT reads 0–50 ppm. Trace OxyScout-T reads 0–1,000 ppm with detection 50 ppm O2*. Percent OxyScout-P reads 0–1 / 0–10 / 0–21 %vol. A 10 ppm tariff example requires Ultra-Trace because Trace detection is 50 ppm.
*Based on controlled lab conditions.
Suppose a hypothetical comparison shows an indication 3 ppm above the measurement standard. The 3 ppm figure illustrates a difference between an indication and a standard. The figure is not an OxyScout accuracy specification.
| Item | Published statement | Not the same as |
|---|---|---|
| Drift | Low single-digit ppm over a typical 1–2 month interval | Detection limit or accuracy |
| Ultra-Trace detection | TBD | The drift statement |
| Trace detection | 50 ppm O2* | A 10 ppm tariff example, which requires Ultra-Trace |
| Hypothetical check | Indication 3 ppm above the standard | An OxyScout specification |
*Based on controlled lab conditions. The 3 ppm row is hypothetical.
Air doesn't damage or poison the OxyScout optical sensor. Once process gas is flowing again, the reading typically recovers within minutes. Electrochemical trace O₂ cells can be used up or permanently damaged by extended air exposure.
NOAA lists dry-air oxygen, excluding water vapor, as 20.946 percent. That figure describes dry air. The recovery statement above is an OxyScout statement and is not taken from NOAA.
A one-point calibration is acceptable. A two-point calibration with zero gas and span gas is recommended.
Drift between calibrations is not the same as detection limit or accuracy. Typically, a 1–2 month interval results in low single-digit ppm of oxygen drift across OxyScout models.
Zero gas is ultra-high-purity (UHP) nitrogen, 99.999%, certified to less than 1 ppm O₂, through a stainless-steel-diaphragm regulator that has been properly bled.
Use a certified span gas near your operating range or critical point. For a 10 ppm application, that is typically 10 ppm O₂ in nitrogen. If a span gas reads far from its certificate value, suspect the gas handling or the cylinder before suspecting the analyzer. An air calibration is often more accurate than a mishandled low-level span gas.
Yes. ScoutView is included with every analyzer and walks through guided zero and span. You can also trigger the same calibrations from an RTU, PLC, or laptop Modbus master. The gas-handling steps are the same either way.