Home/Articles/Calibration

How to calibrate a trace oxygen analyzer in natural gas

How to calibrate a trace oxygen analyzer for natural gas.

Last updated:

Calibration keeps a continuous oxygen analyzer aligned with known gases. For OxyScout™, a two-point calibration with zero gas and span gas is recommended. A one-point calibration is acceptable.

What does calibration mean here?

The International Vocabulary of Metrology defines calibration as an operation that, under specified conditions, establishes a relation between quantity values from measurement standards and the corresponding indications of the instrument. The same entry says calibration should not be confused with adjustment of a measuring system.

For OxyScout, a two-point calibration with zero gas and span gas is recommended. A one-point calibration is acceptable. The steps below are the OxyScout zero and span procedure from the user manual.

How do you calibrate OxyScout?

OxyScout uses a two-point calibration: a zero on oxygen-free gas, then a span on a gas with a known upper point. ScoutView PC software, included with every analyzer, walks through guided zero and span. You can also trigger the same calibrations from an RTU, PLC, or laptop Modbus master. Either method gives the same result. The gas-handling steps below apply to both.

A one-point calibration, often a zero calibration with pure nitrogen, is also acceptable. This is sometimes the path chosen for custody transfer or other applications where any oxygen reading above 0 ppm is unacceptable. This often saves users time and money, since a span gas is not needed.

When should you calibrate?

Calibrate at initial startup, after every Sensor Module replacement, and periodically after that. A typical interval for most installations is 1–2 months. Drift between calibrations also depends on the sample rate and on process quality, as described later on this page.

What gases and equipment do you need?

Zero gas is ultra-high-purity (UHP) nitrogen, 99.999%, certified to less than 1 ppm O₂. Use a stainless-steel-diaphragm regulator. An elastomer-diaphragm regulator will contaminate the gas with atmospheric oxygen and make a true zero impossible.

Span gas should be a certified mixture near your operating range or critical point. For a 10 ppm application, that is typically 10 ppm O₂ in nitrogen. Treat span gas with the same care as zero gas: a stainless-diaphragm regulator, bled thoroughly, stainless tubing, and a leak check. If a span gas reads far from its certificate value, suspect the gas handling or the cylinder before suspecting the analyzer. An air calibration is often more accurate than a mishandled low-level span gas.

A newly installed regulator is full of air on both sides of its diaphragm. Simply flowing gas through it can take days to clear. Bleed the regulator with the steps below before trusting any gas that passes through it.

Why calibrate at service conditions?

Always calibrate with the analyzer venting as it will run in service: the same vent path and the same constant flow setting. Calibrating at one vent condition and operating at another builds a permanent offset into every reading.

Vent to atmosphere is recommended. A return to a process point is acceptable if that point stays at a steady pressure lower than the regulated supply, so gas keeps flowing and the calibration holds.

How do you bleed a newly installed regulator?

  1. Install the regulator on the cylinder and leave its outlet closed.
  2. Crack the cylinder valve open briefly to pressurize the regulator, then close it.
  3. Loosen the regulator-to-cylinder nut until the high-pressure gauge falls to zero, then retighten.
  4. Repeat the pressurize-and-bleed cycle seven to ten times to dilute the air trapped in the regulator body.
  5. Bleed the low-pressure side the same way: pressurize, vent through the outlet until both gauges fall, and repeat seven to ten times.
  6. Leak-check the regulator, gauges, and cylinder valve with liquid leak detector.

Do this for the zero-gas regulator and, later, for the span-gas regulator.

How do you perform a zero calibration?

  1. Flow the UHP nitrogen through the analyzer, using the Sampling System calibration port if one is fitted, or the sample line.
  2. While connecting the calibration line, let gas flow gently through the loose fitting for several seconds before tightening it, so you do not trap a slug of air.
  3. Wait for the reading to fall and fully stabilize. It should be flat for 15–30 minutes, not merely falling slowly. On a first startup, after any opening of the sample system, or with a freshly bled regulator, this can take several hours of continuous purge. Zeroing too early locks residual oxygen in as an offset, which is the most common trace calibration mistake.
  4. Run the zero calibration in ScoutView or over Modbus, then save the settings to permanent memory (Save to Flash).
  5. Return the analyzer to process sample.

How do you perform a span calibration?

  1. Bleed and purge the span-gas regulator the same way as the zero-gas regulator.
  2. Flow the span gas through the analyzer until the reading is stable.
  3. Run the span calibration in ScoutView (guided) or over Modbus. Enter or confirm the temperature, barometric pressure, and humidity values the software or guide calls for, and enter the certified oxygen value.
  4. Save the settings to permanent memory (Save to Flash).
  5. Return the analyzer to process sample and allow the reading to purge back down to process levels.

What does drift between calibrations mean?

Drift between calibrations depends on the sample rate set and on process quality. Typically, a 1–2 month interval results in low single-digit ppm of oxygen drift across OxyScout models. That statement describes drift between calibrations. Drift is not the same as detection limit or accuracy.

How do detection limits differ from drift?

Ultra-Trace OxyScout-UT reads 0–50 ppm. Trace OxyScout-T reads 0–1,000 ppm with detection 50 ppm O2*. Percent OxyScout-P reads 0–1 / 0–10 / 0–21 %vol. A 10 ppm tariff example requires Ultra-Trace because Trace detection is 50 ppm.

*Based on controlled lab conditions.

What would a hypothetical check show?

Suppose a hypothetical comparison shows an indication 3 ppm above the measurement standard. The 3 ppm figure illustrates a difference between an indication and a standard. The figure is not an OxyScout accuracy specification.

ItemPublished statementNot the same as
DriftLow single-digit ppm over a typical 1–2 month intervalDetection limit or accuracy
Ultra-Trace detectionTBDThe drift statement
Trace detection50 ppm O2*A 10 ppm tariff example, which requires Ultra-Trace
Hypothetical checkIndication 3 ppm above the standardAn OxyScout specification

*Based on controlled lab conditions. The 3 ppm row is hypothetical.

How long does recovery take after air?

Air doesn't damage or poison the OxyScout optical sensor. Once process gas is flowing again, the reading typically recovers within minutes. Electrochemical trace O₂ cells can be used up or permanently damaged by extended air exposure.

NOAA lists dry-air oxygen, excluding water vapor, as 20.946 percent. That figure describes dry air. The recovery statement above is an OxyScout statement and is not taken from NOAA.

Sources

FAQ

Frequently asked questions

Is a one-point calibration acceptable on OxyScout?

A one-point calibration is acceptable. A two-point calibration with zero gas and span gas is recommended.

Is drift the same as accuracy?

Drift between calibrations is not the same as detection limit or accuracy. Typically, a 1–2 month interval results in low single-digit ppm of oxygen drift across OxyScout models.

What zero gas does OxyScout use?

Zero gas is ultra-high-purity (UHP) nitrogen, 99.999%, certified to less than 1 ppm O₂, through a stainless-steel-diaphragm regulator that has been properly bled.

What span gas should I use?

Use a certified span gas near your operating range or critical point. For a 10 ppm application, that is typically 10 ppm O₂ in nitrogen. If a span gas reads far from its certificate value, suspect the gas handling or the cylinder before suspecting the analyzer. An air calibration is often more accurate than a mishandled low-level span gas.

Can I calibrate with ScoutView or over Modbus?

Yes. ScoutView is included with every analyzer and walks through guided zero and span. You can also trigger the same calibrations from an RTU, PLC, or laptop Modbus master. The gas-handling steps are the same either way.

Related

Related pages

Talk to an engineer about your application

Book a 30-minute meeting to review your application and we'll send you a quote after the call.

Get a Quote