Products · OxyScout™ by Standard Sensors

Optical Oxygen Analyzers for Natural Gas

OxyScout™ by Standard Sensors is a fixed, continuous optical O2 analyzer for natural gas. The analyzer is an online process oxygen analyzer with a proprietary, fully optical sensor, no electrochemical cells to replace, and simultaneous 4–20 mA and RS-485 Modbus RTU outputs. Models share installation and communications and differ by measuring range.

What is an optical oxygen analyzer for natural gas?

An optical oxygen analyzer measures oxygen with a fully optical sensor rather than an electrochemical (galvanic) cell. OxyScout™ reads trace (ppm) and percent-level O₂ continuously in natural gas and other oil & gas process streams, and sends the result as 4–20 mA or RS-485 Modbus RTU.

Illustrative render of the OxyScout analyzer mounted on a panel at a natural gas site, with piping in the background.
Models

What is the difference between a trace and a percent oxygen analyzer?

A trace oxygen analyzer reports oxygen in parts per million (ppm), and a percent oxygen analyzer reports oxygen in percent by volume (%vol). One percent by volume equals 10,000 ppm. OxyScout Ultra-Trace and OxyScout Trace are the trace models, and OxyScout Percent is the percent model. All three models share the same installation, wiring, firmware and communications, and they differ in measuring range and calibration.

Which OxyScout model fits the application?

ModelMeasuring range*Detection limit*Accuracy*Typical application
OxyScout Ultra-Trace0–50 ppm O₂TBD±2 ppm or ±5% of reading, whichever is greaterSales gas, custody transfer and vapor recovery units with a low-ppm specification
OxyScout Trace0–1,000 ppm O₂50 ppm O₂Typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greaterGathering, vapor recovery units and process
OxyScout Percent0–1, 0–10 or 0–21 %vol O₂, user-selected0.005% O₂ (50 ppm)Typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greaterHigh-level and air-range service, including tank blanketing

*Based on controlled lab conditions.

Choose a range that puts your oxygen limit well inside the scale, with enough headroom above the limit to see an upset rather than a pegged signal. The pipeline oxygen limits guide matches published limits to each range.

Features

Key features

OxyScout combines a proprietary optical sensor, the ZeroLeak™ Cell Block and standard industrial outputs in one analyzer for natural gas service. All three models include the ZeroLeak™ Cell Block.

1

Proprietary optical sensor

OxyScout measures oxygen with a proprietary optical sensor instead of an electrochemical (galvanic) cell. The measurement consumes nothing, and there is no electrolyte to deplete, dry out or leak.

2

No cross-sensitivity or poisoning

OxyScout has no cross-sensitivity or poisoning from H₂S, CO₂ or the other gases that plague galvanic cells.

3

Minimal sample conditioning

Analyzers that need a set sample flow rely on an upstream flowmeter and needle valve to hold it. OxyScout's optical reading does not depend on sample flow rate, so no upstream flowmeter or needle valve is needed. Fewer parts mean fewer things that can fail, and with the ZeroLeak™ Cell Block, fewer parts in the flow path mean fewer leak points.

4

4–20 mA, Modbus and ScoutView

Every model has a 4–20 mA output and RS-485 Modbus RTU for your RTU, PLC or SCADA system, and both outputs work at the same time. ScoutView, the OxyScout configuration software, ships with every analyzer.

5

Engineered for Class 1 Division 2 (Zone 2)

OxyScout is engineered for Class 1 Division 2 (Zone 2) hazardous areas. The IP67 enclosure works at 0 to 50 °C*, and extreme-heat and extreme-cold enclosure options are available.

6

Low cost of ownership

With no electrochemical cells to replace, there are no cell changes to schedule.

*Based on controlled lab conditions.

ZeroLeak™ Cell Block

How does the ZeroLeak™ Cell Block reduce leak points and sample lag?

The ZeroLeak™ Cell Block is the machined block below the analyzer enclosure. The cell block is machined out of one solid block of 6061-T6 aluminum with an electroless nickel plating finish, so minimal fittings are required and there are minimal places for a leak. Because the optical measurement is independent of flow rate, the sample line needs no upstream flowmeter or needle valve, which leaves fewer components between the process tap and the cell block. With fewer components in the flow path, there are fewer joints that could leak and fewer parts that could fail. Sample conditioning is minimal, and an optional sample system is available.

  • One solid block

    The cell block is machined out of one solid block of 6061-T6 aluminum and finished with electroless nickel plating.

  • Minimal fittings, minimal leak points

    Because the flow path is machined into a single block, minimal fittings are required and there are minimal places for a leak.

  • Zero dead legs, minimal volume

    The flow path has zero dead legs and minimal internal volume, so there is less sample lag and a faster response time.

  • Rapid, full purge

    Zero dead legs and minimal internal volume also allow a rapid, full purge of the flow path.

  • Self-draining flow path

    The sample bore slopes down from the inlet on the left to the outlet on the right, so any moisture buildup drains out instead of stagnating in the cell block.

What are the specifications of the analyzer?

The analyzer covers trace and percent-level oxygen with 4–20 mA and Modbus outputs, runs on 9–36 VDC, and is engineered for Class 1 Division 2 (Zone 2) areas. These are typical specifications, and the pressure rating applies without the optional flowmeter installed.

MeasurementThe analyzer uses a fully optical O₂ sensor, so there are no electrochemical cells to replace.
RangesThe Ultra-Trace range is 0–50 ppm, Trace range is 0–1000 ppm and the Percent ranges are 0–1 %, 0–10 % and 0–21 % by volume.
OutputsThe outputs are a 4–20 mA loop (current sink) and RS-485 Modbus RTU (2-wire half-duplex).
PowerThe analyzer runs on 9–36 VDC and draws 35 mA plus the current used by its analog outputs.
Process / ambient temperatureProcess and ambient temperature is 0 to 50 °C, and storage temperature is −20 to 70 °C.
PressureUp to 250 PSI input pressure†
Humidity / flowThe analyzer is specified for non-condensing conditions, and it has no flow restrictions.
Area classificationThe analyzer is engineered for Class 1 Division 2 (Zone 2) hazardous areas.
Size / weightThe enclosure is 9 in high, 7 in wide and 2.5 in deep, and the analyzer weighs 4 lb.
Process connectionThe process connection is 1/8 in tube compression.
OptionsThe options include a flowmeter to regulate vent flow to atmosphere if required, a sample system (pressure regulator and gauge, secondary filter, and a 3-way valve for calibration gas or process sample), and an extreme-heat or extreme-cold enclosure.
Response time*Typical response time is under 10 seconds.
Detection limit*The limit of detection is TBD on OxyScout Ultra-Trace (0–50 ppm), 50 ppm O₂ on OxyScout Trace (0–1,000 ppm) and 0.005% O₂ (50 ppm) on OxyScout Percent (0–1, 0–10 and 0–21 %vol).
Accuracy*Accuracy is ±2 ppm or ±5% of reading, whichever is greater on OxyScout Ultra-Trace. On OxyScout Trace and on all OxyScout Percent ranges, accuracy is typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greater.
Resolution*Resolution on OxyScout Trace and OxyScout Percent is 0.002% O₂ at 0.2% O₂.

*Based on controlled lab conditions. Standard Sensors Corporation reserves the right to make revisions and/or changes without notice.

†Without Flowmeter installed. Higher ratings if regulator installed upstream.

Drawing of the analyzer enclosure with a hand for scale, labelled 9 in high and 7 in wide.

How is the analyzer installed and connected?

The sample tap uses 1/8 in tube compression. The analyzer measures 9 × 7 × 2.5 in, weighs 4 lb and runs on 9–36 VDC at 35 mA plus the current used by its analog outputs. Wire the 4–20 mA loop or RS-485 Modbus RTU to your RTU, PLC or SCADA, and set your own alarm points.

Without the optional flowmeter installed, the analyzer is rated up to 250 psi. Higher ratings are available if a regulator is installed upstream. The optional sample system adds a pressure regulator and gauge, a secondary filter and a 3-way valve for calibration gas or process sample.

What options and accessories are available?

  • The flowmeter regulates vent flow to atmosphere, if required.
  • The sample system adds a pressure regulator and gauge, a secondary filter, and a 3-way valve to select calibration gas or process sample.
  • The extreme-heat or extreme-cold enclosure gives the best stability in direct sunlight or below-freezing conditions.

How does it compare with an electrochemical analyzer?

Electrochemical (galvanic) cellOptical (OxyScout)
What wears outThe cell itself wears out and is replaced on a schedule.No electrochemical cell has to be replaced.
H₂S, CO₂ and other gasesThese gases can affect the cell.OxyScout has no cross-sensitivity or poisoning.
OutputsThe outputs vary by model.The outputs are 4–20 mA and RS-485 Modbus RTU.

For the full comparison, read optical vs electrochemical oxygen analyzers.

Total cost of ownership
  • OxyScout has no electrochemical cells to replace, so there are no cell changes to schedule.

Where is the analyzer used?

Diagram of a natural gas line with an OxyScout analyzer sampling the gas and sending a continuous oxygen reading to a control system.
Continuous oxygen measurement

OxyScout measures oxygen continuously in natural gas, and the reading goes out over 4–20 mA and RS-485 Modbus RTU at the same time.

The analyzer uses a fully optical O₂ sensor, so there are no electrochemical cells to replace.

FAQ

Frequently asked questions

What does an optical oxygen analyzer measure?

An optical oxygen analyzer measures oxygen (O₂) in a gas sample, continuously, with a fully optical sensor. OxyScout covers trace ranges of 0–50 ppm and 0–1000 ppm by volume and percent ranges of 0–1 %, 0–10 % and 0–21 % by volume.

What is the difference between OxyScout Trace and OxyScout Percent?

OxyScout Trace reads 0–1,000 ppm O₂ with a 50 ppm detection limit, and OxyScout Percent reads 0–1, 0–10 or 0–21 %vol O₂ with a 0.005% O₂ (50 ppm) detection limit.*

What are the outputs and power requirements?

The outputs are a 4–20 mA loop (current sink) and RS-485 Modbus RTU (2-wire half-duplex). Power is 9–36 VDC with a 35 mA draw plus the current used by the analog outputs.

What hazardous areas and environments is it designed for?

The analyzer is engineered for Class 1 Division 2 (Zone 2) hazardous areas and has an IP67 enclosure. Process and ambient temperature is 0 to 50 °C. The analyzer is rated up to 250 psi without the flowmeter installed. Higher ratings are available if a regulator is installed upstream. Extreme-heat and extreme-cold enclosures are available.

How fast does it respond, and how low can it detect?

Typical response time is under 10 seconds.* The limit of detection is 50 ppm O₂ on OxyScout Trace and 0.005% O₂ (50 ppm) on OxyScout Percent.* *Based on controlled lab conditions.

Does H₂S or CO₂ in the gas affect it?

OxyScout has no cross-sensitivity or poisoning from H₂S, CO₂ or the other gases that plague galvanic cells.

How do I get a quote?

Book a free 30-minute online meeting to review your application, range and quantities, then we send a quote. You can also call (936) 259-4643 or email [email protected].

Does H2S affect OxyScout?

No. The optical sensor has no cross-sensitivity to H2S. In field testing of the sensing technology, the sensor ran in gas containing high H2S concentration (percent level) with no effect. OxyScout is intended for dry, non-condensing sample gas that contains H2S. For wet sour gas, contact Standard Sensors to review the application suitability.

Does CO2 affect OxyScout?

No. The sensing technology has been field tested in up to 100% CO2 with no effect on the reading and no damage.

What can interfere with the sensor?

Organic solvent vapors, chlorine gas, and nitrogen dioxide (NO2) interfere with the sensor. Methane, CO2, and H2S do not interfere.

What happens if liquid water reaches the sensor?

Liquid water offsets the reading temporarily. Once the sample path is purged, the reading recovers. OxyScout is specified for non-condensing sample conditions.

How long does the optical sensor last?

The sensor is covered by the 12-month warranty. Standard Sensors expects more than one year of service, and longer in clean, dry sample conditions. Long-term life data is still being collected.

How is OxyScout calibrated?

A two-point calibration with zero gas and span gas is recommended. A one-point calibration is acceptable. Drift between calibrations depends on the sample rate set and on process quality. Typically, a 1–2 month interval results in low single-digit ppm of oxygen drift across OxyScout models. Drift between calibrations is not the same as detection limit or accuracy.

How fast does OxyScout recover after exposure to air?

Air doesn't damage or poison the OxyScout optical sensor. Once process gas is flowing again, the reading typically recovers within minutes. Electrochemical trace O₂ cells can be used up or permanently damaged by extended air exposure.

Where is OxyScout built, and what are the warranty and lead time?

OxyScout is built in the USA. The standard warranty is 12 months from shipment and includes the sensor. Lead time can be less than one week.

Is OxyScout a personal oxygen monitor or portable gas detector?

No. OxyScout is a fixed process analyzer that samples natural gas from a pipe. OxyScout is not for personal safety, medical, combustion or dissolved-oxygen use.

Gas compatibility for OxyScout

Gas or conditionEffect on OxyScout
Natural gas, methane (CH4)No cross-sensitivity
Hydrogen sulfide (H2S)No cross-sensitivity of the optical sensor. In field testing of the sensing technology, high H2S concentration (percent level) had no effect. Intended for dry, non-condensing sample gas that contains H2S. Wet sour gas requires application review.
Carbon dioxide (CO2)No cross-sensitivity. Field tested up to 100%.
Liquid waterTemporary offset. Recovers after purging.
Other organic solvent vaporsInterferes
Chlorine gas, nitrogen dioxide (NO2)Interferes
Models

OxyScout models, SKU and MPN

SKU and MPN are the same value for each model. Detection limits below use the lab footnote.

ModelSKUMPNMeasuring range
OxyScout Ultra-TraceOxyScout-UTOxyScout-UT0–50 ppm
OxyScout TraceOxyScout-TOxyScout-T0–1,000 ppm
OxyScout PercentOxyScout-POxyScout-P0–1 / 0–10 / 0–21 %vol

Detection limits from the published specifications: Trace 50 ppm*; Percent 0.005% O2 (50 ppm)*. Trace is not the match for a 10 ppm tariff limit, because the published detection limit is 50 ppm.

*Based on controlled lab conditions.